Medical Device Simulators

Medical Device Simulators · Interactive Lab

Magnetic Resonance Imaging

Change the main magnetic field B₀ to observe the Larmor resonance frequency of hydrogen protons — the frequency an MRI scanner must transmit and receive at.

Parameters

Use the controls or enter precise values.

0.2 T7 T
0.2 T7 T

Why these controls?

Field strength is the main dial of MRI physics: the Larmor frequency f₀ = (γ/2π)·B₀ grows in direct proportion to B₀. Move it to see how 1.5 T and 3 T systems differ, and use the comparison field to put two field strengths side by side.

Calculated result

Larmor resonance frequency

Resonance frequency f₀

63.87

MHz

ω₀ = 2πf₀ = 4.0131e+8 rad/s

Field comparison

1.5 T

63.87 MHz

3.0 T

127.74 MHz

Frequency ratio: ×2.00 — equals the field ratio, because f₀ ∝ B₀.

Frequency f₀
63.87 MHz
Angular frequency ω₀
401.3 Mrad/s

f₀ is in megahertz (cycles per second); ω₀ is in radians per second. They differ by a factor of 2π — do not mix the two units.

Frequency vs. magnetic field

Linear through the origin — slope γ/2π ≈ 42.58 MHz/T

0.001.753.505.257.000751492242981.5 T3.0 TMagnetic field B₀ (T)Frequency f₀ (MHz)

Formula

f₀ = (γ/2π) · B₀
  • f₀ = Larmor frequency in megahertz (MHz)
  • B₀ = main magnetic field strength in tesla (T)
  • γ/2π = gyromagnetic ratio of ¹H ≈ 42.58 MHz/T
  • ω₀ = 2πf₀ = angular frequency in radians per second (rad/s)

Scope & assumptions

  • This module shows only the resonance relationship f₀ = (γ/2π)B₀ for hydrogen protons — it does NOT produce an MRI image.
  • The field is assumed perfectly uniform; real scanners have shim imperfections and gradient fields for spatial encoding.
  • Chemical shift, relaxation (T1/T2), and RF pulse effects are not modelled.
  • Other nuclei (¹³C, ³¹P, ²³Na) have different gyromagnetic ratios and resonate at different frequencies.