Medical Device Simulators

Medical Laser

Separate optical power, energy, exposure time, and circular beam area in an ideal homogeneous-beam model.

Medical Device Simulators

Operating mode

Beam inputs

Why these controls?

Diameter sets the beam area A = πr²; power and time (or pulse energy and duration) set the energy. Move them to see how the same energy spread over a larger area lowers irradiance and fluence.

Calculated distribution

Homogeneous circular beam

Beam area

0.503 cm²

Total energy

10.000 J

Average irradiance

9.947 W/cm²

Energy density (fluence)

19.894 J/cm²

Exposure

2.0 s

Diameter: 8.0 mm → radius: d/2 = 4.00 mm

Unit conversion: 4.00 mm ÷ 10 = 0.400 cm

Area: A = πr² = π × (0.400 cm)² = 0.503 cm²

Energy: E = Pt = 5.0 W × 2.0 s = 10.000 J

Spatial distribution

Same energy, different diameter

8.0 mm16.0 mmH = 19.89 J/cm²H = 4.97 J/cm²diameter ×2 → area ×4 → fluence ÷4
Lower fluenceMediumHigher fluence

Color represents relative calculated fluence only. It is not a tissue temperature, damage, depth, or clinical effectiveness map.

Diameter response

Fluence vs. beam diameter

0.0318.3636.6954.91273.2Beam diameter (mm)Fluence (J/cm²)
Energy is held fixed at 10.00 J.

Model equations

I = P/A W/cm²

H = E/A J/cm²

E = Pt continuous beam

Student checks

  • Double diameter and confirm that area becomes four times larger.
  • At fixed energy, confirm that doubled diameter reduces fluence to one quarter.
  • In pulse mode, keep energy fixed and change duration: fluence stays constant while average power changes.
  • Investigate why homogeneous and Gaussian beam profiles differ in real lasers.